21. Push OK Add Measurement on the side menu, and then Menu Off on the front panel. The mean value should appear in
a measurement p
ane at the bottom of the display.
22. Enter the meas
ured value in the test record. (See page 47, DC Offset Accuracy Tests.)
23. Repeat the pro
cedure (steps 6, 7, 8 and 22) for each v olts/division setting shown in the test record.
24. Repeat all ste
ps, starting with step 1, for each oscilloscope channel you want to check.
This comp let
es the procedure.
84 MDO3000 Series Specications and Performance Verication