Commands Listed in Alphabetical Order
Conditions
Available for M
SO/DPO4000B series models, 2-channel MDO3000 models, and
MDO4000C models without option SA3 or SA6 only.
Group
Vertical
Syntax
AUXin:PRObe:RESistance?
Examples
AUXin:PRObe:RESistance? might return :AUXin:PROBE:RESISTANCE
1.0000E+6
indicat
ing that the input resistance of the probe attached to the front
panel Aux In connector is 1 MΩ.
NOTE. This query will return 0.0 if no probe is attached or the attached probe
does not report the input resistance.
AUXin:PRObe:SIGnal
This command ch
anges the input bypass setting on VPI probes that support input
bypass, for example the TCP0001. If sent to a probe that does not support input
bypass, it is ignored.
Conditions
Available for MSO/DPO4000B series models, 2-channel MDO3000 models, and
MDO4000C models without option SA3 or SA6 only.
Group
Vertical
Syntax
AUXin:PRObe:SIGnal {BYPass|PASS}
AUXin:PRObe:SIG
nal?
Arguments
ByPass sets the probe to Bypass mode.
PASS sets the probe to Pass mode.
AUXin:PRObe:UNIts? (Query Only)
Returns a string describing the units of measure of the probe attached to theAux
Input connector.
Conditions
Available for MSO/DPO4000B series models, 2-channel MDO3000 models, and
MDO4000C models without option SA3 or SA6 only.
MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual 2-155