Commands Listed in Alphabetical Order
Arguments
CH1–CH4 specifi
es an analog channel as the bit source waveform.
D0-D15 specifies a digital channel as the bit source waveform.
(MSO/MDO4000
/B models only as well as MDO3000 and MDO400C models
with option MDO3MSO or MDO4MSO installed.)
BUS:B<x>:PARallel:CLOCk:EDGE
Specifies the clock edge for the parallel bus.
Conditions
Requires an MSO/MDO model, with the MDO3MSO option installed for the
MDO3000 series. The MSO4000B and MDO4000/B/C models can trigger on
parallel buses without an application module.
Group
Bus
Syntax
BUS:B<x>:PARallel:CLOCk:EDGE {EITher|RISing|FAL ling}
BUS:B<x>:PARallel:CLOCk:EDGE?
Arguments
EIther specifies either edge as the clock edge.
RISing specifies the rising edge as the clock edge.
FALling specifies the falling edge as the clock edge.
BUS:B<x>:PARallel:CLOCk:ISCLOCKed
Specifies the state of the clock function for the parallel bus.
Conditions
Requires an MSO/MDO model, with the MDO3MSO option installed for the
MDO3000 series. The MSO4000B and MDO4000/B/C models can trigger on
parallel buses without an application module.
Group
Bus
Syntax
BUS:B<x>:PARallel:CLOCk:ISCLOCKed {YES|NO}
BUS:B<x>:PARallel:CLOCk:ISCLOCKed?
Arguments
YES specifes that the parallel bus is clocked.
NO specifes that the parallel bus is not clocked.
MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual 2-189