Commands Listed in Alphabetical Order
CALibrate:FA
Ctory:STATus:RF? (Query Only)
Returns the factory calibration status value saved in nonvolatile memory for the
RF portion of the oscilloscope.
This query is synonymous with CALibrate:RESults:FACtory:RF?.
Conditions
MDO models only
Group
Calibration and Diagnostic
Syntax
CALibrate:FACtory:STATus:RF?
Returns
PASS: shows the RF portion of the factory calibration has succeeded.
INIT: shows the RF portion of the factory calibration has not been adjusted.
RUNNING: shows the RF portion of the factory calibration is in progress.
Examples
CAL:FAC:STAT:RF? might return RUNNING, indicating that the factory
calibration for the RF portion of the instrument is in progress.
CALibrate:FACtory:STATus:SCOPE? (Query Only)
Returns the factory calibration status value saved in nonvolatile memory
for the non-RF portion of the oscilloscope. It is synonymous with the
:CALibrate:RESults:FACtory:SCOPE? query.
Conditions
MDO models only
Group
Calibration and Diagnostic
Syntax
CALibrate:FACtory:STATus:SCOPE?
Returns
PASS: shows the non-RF portion of the factory calibration has succeeded.
INIT: shows the non-RF portion of the factory calibration has not been adjusted.
RUNNING: shows the non-RF portion of the factory calibration is in progress.
FAIL: show the non-RF portion of the factory calibration has failed.
MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual 2-211