Commands Listed in Alphabetical Order
<NR1> = 0 indica
tes the calibration completed without errors.
Examples
*CAL? starts the internal signal path calibration and might return 0 to indicate that
the calibration was successful.
CALibrate:FACtory:STATus? (Query Only)
Returns the factory calibration status value saved in nonvolatile memory.
Group
Calibra
tion and Diagnostic
Syntax
CALibrate:FACtory:STATus?
Examp
les
CAL:FAC:STAT? might return CALIBRATE:FACTORY:STATUS PASS indicating
that factory calibration passed.
CALibrate:FACtory:STATus:AFG? (Query Only)
Thi
s query returns the factory calibration status for the Arbitrary Function
Generator portion of the instrument, if present.
Th
is query is synonymous with CALibrate:RESults:FACtory:AFG?
Group
Calibration and Diagnostic
Syntax
CALibrate:FACtory:STATus:AFG?
Related Commands
CALibrate:FACtory:STATus:SCOPE?, CALibrate:FACtory:STATus:RF?
Returns
PASS - factory calibration has passed
INIT - factory calibration is unadjusted
RUNNING - factory calibration is running
Examples
CALIBRATE:FACTORY:STATUS:AFG? might return RUNNING, indicating that the
factory calibration for the AFG portion of the instrument is in progress.
2-210 MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual

Table of Contents

Other manuals for Tektronix MDO3000 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix MDO3000 Series and is the answer not in the manual?

Tektronix MDO3000 Series Specifications

General IconGeneral
BrandTektronix
ModelMDO3000 Series
CategoryTest Equipment
LanguageEnglish

Related product manuals