Commands Listed in Alphabetical Order
Syntax
CH<x>:PRObe:P
ROPDELay <NR3>
CH<x>:PRObe:PROPDELay?
Arguments
<NR3> is a floating point number that specifies the propagation time delay for the
connected probe.
Examples
CH1:PROBE:PROPDELAY 100E-12 sets the CH1 propagation delay to 100 ps.
CH1:PROBE:PROPDELAY? might return :CH1:PROBE:PROPDELAY 1.1E-9
indicating that the CH1 propagation delay is set to 1.1 ns.
CH<x>:PRObe:RECDESkew? (Query Only)
Returns the recommended deskew for the probe connected to the specified
channel.
Group
Vertical
Syntax
CH<x>:PRObe:RECDESkew?
CH<x>:PRObe:RESistance? (Query Only)
Returns the resistance factor of the probe attached to channel <x>, where x is the
channel number.
Group
Vertical
Syntax
CH<x>:PRObe:RESistance?
Examples
CH2:PRObe:RESistance? might return :CH2:PROBE:RESISTANC E
1.0000E+6
indicating that the input resistance of the probe attached to Channel 2
is 1 MΩ.
NOTE. This query will return 0.0 if no probe is attached or the attached probe
does not report the input resistance.
2-232 MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual