Commands Listed in Alphabetical Order
MASK:TESt:COM
PLetion:CRITerion
Arguments
ON turns on this features, so that the instrument stops acquiring data upon failure.
OFF turns off this feature.
<NR1> is an integer. 0 turns off this feature; any other value turns it on.
Examples
MASK:TESt:STOP:FAILure 1 stops data acquisition upon test failure.
MASK:TESt:STOP:FAILure? might return 0, indicating the instrument is not
set to stop acquiring data upon test failure.
MASK:TESt:THReshold
This command
specifies the number of failed tested waveforms needed in a
pass/fail mask test to cause the test status to change to ‘Failing’.
Aseriesofe
xamples showing how to use mask commands for typical tasks
is included in an appendix. (See page E-1, Mask/Limit Command Sequence
Examples.)
Conditions
This comman
d requires a DPO4LMT application module for 4000 series models,
and an MDO3LMT application module for MDO3000 series models.
Group
Mask
Syntax
MASK:TESt:THReshold {<NR1>|INFInite}
MASK:TESt:THReshold?
Related Commands
MASK:TESt:STATE,
MASK:TESt:COMPLetion:CRITerion
Arguments
<NR1> is an integer that specifies the number of tested waveform violations
occurring in each mask test that will change the test status to ‘Failing’. The
maximum number of failed tested waveforms that can be specified is 1,000,000.
The default is 1.
INFInite indicates that there is no limit to the number of failed tested waveforms
that will change a test status to ‘Failing’.
MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual 2-395