Commands Listed in Alphabetical Order
MEASUrement:
MEAS<x>:MEAN? (Query Only)
Returns the mean value accumulated for this measurement since the last statistical
reset. Measurements are specified by x, which ranges from 1 through 8.
Group
Measurement
Syntax
MEASUrement:MEAS<x>:MEAN?
Examples
MEASUREMENT:MEAS1:MEAN? might return :MEASUREMENT:MEA S1:MEAN
514.71E-09.
MEASUrement:MEAS<x>:MINImum? (Query Only)
Returns the minimum value for this measurement since the last statistical reset.
Measurements are specified by <x>, which ranges from 1 through 8.
Group
Measurement
Syntax
MEASUrement:MEAS<x>:MINImum?
Examples
MEASUREMENT:MEAS1:MINIMUM? might return
:MEASUREMENT:MEAS1:MINIMUM 1.75E-09.
MEASUrement:MEAS<x>:SOUrce<x>
For SOURce1: This command specifies the source for all single channel
measurements. For delay or phase measurements, This command specifies the
waveform to measure "from". This is equiv
alent to setting the "From:" waveform
in the "Measure Delay" side menu or the "Measure Phase" side menu. SOUrce is
equivalent to SOURCE1.
For SOUrce2: This command specifies the waveform to measure "to" when taking
a delay measurement or phase measurement. This is equivalent to setting the"To:"
waveform in the "Measure Delay" side menu or the "Measure Phase" side menu.
Measurements are specified by <x>, which ranges from 1 to 8.
NOTE. If you d o not specify a numeric sufx,thesourceisassumedtobeSOUrce1.