Commands Listed in Alphabetical Order
Arguments
LESSthan sets t
he data qualifier to less than.
MOREthan sets the data qualifier to greater than.
EQual sets the data qualifier to equal.
UNEQual sets the data qualifier to not equal.
LESSEQual sets the data qualifier to less than or equal.
MOREEQual sets the data qualifier to greater than or equal.
INrange sets the data qualifier to in range.
OUTrange s
ets the data qualifier to out of range.
Examples
SEARCH:SEARCH1:TRIGGER:A:BUS:B1:FLEXRA Y:DATA:QUALIFIER
LESSTHAN
sets the data qualifier to less than.
SEARCH:SEARCH1:TRIGGER:A:BUS:B1:FLEXRA Y:DATA:QUALIFIER? might
return
SEARCH:SEARCH1:TRIGGER:A:BUS:B1:FLEXRAY:D ATA:QUALIFIER
EQUAL
indicating the data qualifier is =.
SEARCH:SEARCH<x>:TRIGger:A:BUS:B<x>:FLEXray:DATa:SIZe
This command specifies the length of the data string, in bytes, to use when
searching on the Fl
exRay bus data field. The search condition needs to be set to
ID or IDANDDATA (using SEARCH:SEARCH<x>:TRIGger:A:BUS:B<x>:
FLEXray:CONDition).
SEARCH<x> is the search number, which is always 1,
and
B<x>
Conditions
This command requires a DPO4AUTOMAX application module for
DPO/MSO4000B and MDO4000/B/C models or MDO3FLEX application module
for MDO3000 models.
Group
Search
Syntax
SEARCH:SEARCH<x>:TRIGger:A:BUS:B<x>:FL EXray:DATa:SIZe <NR1>
SEARCH:SEARCH<x>:TRIGger:A:BUS:B<x>:FL EXray:DATa:SIZe?
Arguments
<NR1> is the length of the data string in bytes. Length range is 1 to 8.
Examples
SEARCH:SEARCH1:TRIGGER:A:BUS:B1:FLEXRA Y:DATA:SIZE 8 sets the
data size to 8 bytes.
MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual 2-653