Commands Listed in Alphabetical Order
SEARCH:SEARCH
1:TRIGGER:A:BUS:B1:FLEXRAY:FRAMEID:HIVALU E
"XXXXXXXXXXX"
indicating the frame id high value is don't cares.
SEARCH:SEARCH<x>:TRIGger:A:BUS:B<x>:FLEXray:FRAMEID:QUALifier
This command specifies the qualifier to use when searching on the FlexRay
bus frame ID field. The search condition needs to be set to IDentifier (using
SEARCH:SEARCH<x>:TRIGger:A:BUS:B<x>:FLEXray:CONDition).
SEARCH<x> is the search number, which is always 1, and B<x>
Conditions
This command requires a DPO4AUTOMAX application module for
DPO/MSO4000B and MDO4000/B/C models or MDO3FLEX application module
for MDO3000 models.
Group
Search
Syntax
SEARCH:SEARCH<x>:TRIGger:A:BUS:
B<x>:FLEXray:FRAMEID:QUALifier
{LESSthan|MOREthan|EQual|UNEQual|LESSE Qual|MOREEQual|
INrange|OUTrange}
SEARCH:SEARCH<x>:TRIGger:A:BUS:B<x>:FL EXray:FRAMEID:
QUALifier?
Arguments
LESSthan sets the frame ID qualifier to less than.
MOREthan sets the frame ID qualifier to more than.
EQual sets the frame ID qualifier to equal.
UNEQual sets the frame ID qualifier to unequal
LESSEQual sets the frame ID qualifier to less than or equal.
MOREEQual sets the frame ID qualifier to greater than or equal.
INrange sets the frame ID qualifier to in range.
OUTrange sets the frame ID qualifier to outside of range.
Examples
SEARCH:SEARCH1:TRIGGER:A:BUS:B1:FLEXRA Y:FRAMEID:QUALIFIER
OUTRANGE
sets the frame ID qualifier to out of range.
SEARCH:SEARCH1:TRIGGER:A:BUS:B1:FLEXRA Y:FRAMEID:QUALIFIER?
might return
SEARCH:SEARCH1:TRIGGER:A:BUS:B1:FLEXRA Y:FRAMEID:QUALIFIER
EQUAL
indicating the frame ID qualifier is EQUAL.