EasyManuals Logo

Tektronix MDO3000 Series Program Manual

Tektronix MDO3000 Series
1158 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #718 background imageLoading...
Page #718 background image
Commands Listed in Alphabetical Order
SEARCH:SEARC
H<x>:TRIGger:A:BUS:B<x>:MIL1553B:STATu s:BIT:INSTR
When the MIL-STD-1553 bus search condition is set to ST ATus, this command
specifies the status word instrumentation bit value (bit 10) to be used in thesearch.
Returned val
ues are 0, 1, or X (don’t care, which is the default).
SEARCH<x> is
the search number, which is always 1, and
B<x>
Conditions
This command requires a DPO4AERO application module for DPO/MSO4000B
and MDO4000/B/C models or MDO3AERO application module for
MDO3000 models.
Group
Search
Syntax
SEARCH:SEARCH<x>:TRIGger:A:BUS:B<x>:MI L1553B:STATus:BIT:
INSTR { 0|1|X|ZERo|ONE|NOCARE|OFF|ON }
SEARCH:SEARCH<x>:TRIGger:A:BUS:B<x>:MI L1553B:STATus:BIT:
INSTR?
Related Commands
SEARCH:SEARCH<x>:TRIGger:A:BUS:B<x>:MIL1553B:CONDition
Arguments
0
1
X
sets the value to X (“don’t care”) which is the default.
ZERO sets the value to 0.
ONE sets the value to 1.
NOCARE sets the value to X (“don’t care”) which is the default.
OFF sets the value to 0.
ON sets the value to 1.
Examples
SEARCH:SEARCH1:TRIGger:A:BUS:B1:MIL155 3B:STATus:BIT:INSTR
NOCARE
sets the status word status word instrumentation bit value to be used in
the search to X.
SEARCH:SEARCH1:TRIGger:A:BUS:B1:MIL155 3B:STATus:BIT:INSTR?
might return 1.
2-688 MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual

Table of Contents

Other manuals for Tektronix MDO3000 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix MDO3000 Series and is the answer not in the manual?

Tektronix MDO3000 Series Specifications

General IconGeneral
BrandTektronix
ModelMDO3000 Series
CategoryTest Equipment
LanguageEnglish

Related product manuals