Commands Listed in Alphabetical Order
SEARCH:SEARCH
1:TRIGger:A:BUS:B1:MIL1553B:TIMe:LESSLim it?
might return 2.0000e-6, indicating that the RT or IMG has been set to 2.00 μS.
SEARCH:SEARCH<x>:TRIGger:A:BUS:B<x>:MIL1553B:TIMe:MORELimit
When the MIL-STD-1553 bus search condition is set to TIMe, this command
specifies either the maximum remote terminal response time (RT) limit
for the amount of time the terminal has to transmit, or it specifies the
maximum in
ter-message gap (IMG). (You can specify RT or IMG using the
SEARCH:SEARCH<x>:TRIGger:A:BUS:B<x>:MIL1553B:CONDition TIMe
command.)
SEARCH<x> is the search number, which is always 1, and B<x>
Conditions
This command requires a DPO4AERO application module for DPO/MSO4000B
and MDO4000/B/C models or MDO3AERO application module for
MDO3000 mod
els.
Group
Search
Syntax
SEARCH:SEARCH<x>:TRIGger:A:BUS:B<x>:MI L1553B:TIMe:MORELimit
<NR3>
SEARCH:SEARCH<x>:TRIGger:A:BUS:B<x>:MI L1553B:TIMe:MORELimit?
Related Commands
SEARCH:SEARCH<x>:TRIGger:A:BUS:B<x>:MIL1553B:CONDition
SEARCH:SEARCH<x>:TRIGger:A:BUS:B<x>:MIL1553B:TIMe:LESSLimit
Arguments
<NR3> is a floating point number that specifies either the maximum remote
terminal response time (RT) or the inter-message gap (IMG) in seconds.
Examples
SEARCH:SEARCH1:TRIGger:A:BUS:B1:MI
L1553B:TIMe:MORELimit
80.0000e-6
would set either the remote terminal response time (RT) or the
inter-message gap (IMG) to be used in the search to 80.0 μS.
SEARCH:SEARCH1:TRIGger:A:BUS:B1:MIL155 3B:TIMe:MORELimit?
might return 100.0000e-6, indicating that the RT or IMG has been set to 100.00 μS.
SEARCH:SEARCH<x>:TRIGger:A:BUS:
B<x>:MIL1553B:TIMe:QUALier
When the MIL-STD-1553 bus search condition is set to TIMe, this command
specifies the search data time qualifier. (This includes a smaller set of arguments
than other qualifier commands.)
B<
x>
2-694 MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual

Table of Contents

Other manuals for Tektronix MDO3000 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix MDO3000 Series and is the answer not in the manual?

Tektronix MDO3000 Series Specifications

General IconGeneral
BrandTektronix
ModelMDO3000 Series
CategoryTest Equipment
LanguageEnglish

Related product manuals