Commands Listed in Alphabetical Order
UNEQual search
es for pulse widths unequal to the time set using
SEARCH:SEARCH<x>:TRIGger:A:PULSEWidth:WIDth within a tolerance
of ±5%.
WIThin searches for pulse widths less than the specified HIGHLimit
and greater than the specified LOWLimit. The limits are specified
using SEARCH:SEARCH<x>:TRIGger:A:PULSEWidth:LOWLimit and
SEARCH:SEARCH<x>:TRIGger:A:PULSEWidth:HIGHLimit.
OUTside searches for pulse widths greater than the specified HIGHLimit
or less than the specified LOWLimit. The limits are specified using
SEARCH:SEARCH<x>:TRIGger:A:PULSEWidth:LOWLimit and
SEARCH:S
EARCH<x>:TRIGger:A:PULSEWidth:HIGHLimit.
Examples
SEARCH:SEARCH<x>:TRIGger:A:PULSEWidth: WHEn LESSTHAN
2.0000E-9
causes the oscilloscope to search the record for pulses shorter than
2 ns duration.
SEARCH:SEARCH<x>:TRIGger:A:PULSEWidth: WHEn? might return
SEARCH:SEARCH<x>:TRIGGER:A:PULSEWIDTH: WHEN GREATER THAN
2.0000E-9
indicating that searches are set to detect pulses greater than 2 ns
duration.
To find all pulses in the waveform record with durations (widths) that fall outside
of the range of 100 nanoseconds to 110 nanoseconds:
SEARCH:SEARCH1:TRIGger:A:PULSEWidth:LO WLimit 100.0E-9
SEARCH:SEARCH1:TRIGger:A:PULSEWidth:HI GHLimit 110.0E-9
SEARCH:SEARCH1:TRIGger:A:PULSEWidth:WH En OUTside
SEARCH:SEARCH<x>:TRIGger:A:PULSEWidth:WIDth
This command specifies the width setting to use, in seconds, when searching
the waveform record for pulses of a certain width (duration).
SEARCH<x> is
the search number, which is always 1.
Group
Search
Syntax
SEARCH:SEARCH<x>:TRIGger:A:PULSEWidth: WIDth <NR3>
SEARCH:SEARCH<x>:TRIGger:A:PULSEWidth: WIDth?
Arguments
<NR3> is a floating point number that specifies the pulse width.
2-738 MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual