Commands Listed in Alphabetical Order
SEARCH:SEARC
H<x>:TRIGger:A:SETHold:SETTime
This command specifies the setup time setting for an setup/hold search to
determine where to place a mark.
SEARCH<x> is the search number, which is
always 1.
Group
Search
Syntax
SEARCH:SEARCH<x>:TRIGger:A:SETHold:SET Time <NR3>
SEARCH:SEARCH<x>:TRIGger:A:SETHold:SET Time?
Arguments
<NR3> is a floating point number that specifies the setup time for setup and hold
violation searching.
SEARCH:SEARCH<x>:TRIGger:A:SETHold:THReshold{:MATH|:MATH1}
This command specifies the search setup and hold threshold for the math
waveform. This helps to determine where to place search marks. Searc
h<x> is
the search number, which is always 1.
Conditions
MSO/MDO4000/B models only as well as MDO3000 and MDO400C models
with option MDO3MSO or MDO4MSO installed.
Group
Search
Syntax
SEARCH:SEARCH<x>:TRIGger:A:SETHold:THR eshold{:MATH|:MATH1}
<NR3>
SEARCH:SEARCH<x>:TRIGger:A:SETHold:THR eshold{:MATH|:MATH1}?
Arguments
<NR3> is a floating point number that specifies the lower threshold in volts.
SEARCH:SEARCH<x>:TRIGger:A:SETHold:THReshold:REF<x>
This command specifies the search setup and hold threshold for the selected
reference waveform. This helps to determine where to place search marks.
Search<x> is the search number, which is always 1. REF<x> is the reference
waveform number, which can be 1–4.
Group
Search
2-744 MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual