Commands Listed in Alphabetical Order
Arguments
<NR1> is the dat
aoffsetvalue.
TRIGger:A:BUS:B<x>:AUDio:DATa:QUALifier
This command sets the qualifier (<, >, =, <=, >=, not =, in range, out of range) to
be used when t
riggering on an audio bus signal. The trigger condition must be set
to DATA using TRIGger:A:BUS:B<x>:AUDio:CONDition.
B<x> is the b
us number (1-4) 1-3 for the MDO4000C and 1-2 for the MDO3000.
Conditions
This command requires a DPO4AUDIO application module on DPO/MSO4000B
and MDO4000/B/C models and a MDO3AUDIO application module on
MDO3000 models.
Group
Trigger
Syntax
TRIGger:A:BUS:B<x>:AUDio:DATa:QUALifie r
{LESSthan|MOREthan|EQual|UNEQual|LESSE Qual|MOREEQual|INrange|OUTrange}
TRIGger:A:BUS:B<x>:AUDio:DATa:QUALifie r?
Arguments
LESSthan sets the qualifier to less than.
MOREthan sets the qualifier to greater than.
EQual sets the qualifier to equal.
UNEQual sets the qualifier to not equal.
LESSEQual sets the qualifier to less than or equal.
MOREEQual sets the qualifier to greater than or equal.
INrange sets the qualifier to in range.
OUTrange sets the qualifier to out of range.
TRIGger:A:BUS:B<x>:AUDio:DATa:VALue
This command sets the lower word valuetobeusedwhentriggeringon
an audio bus signal. The trigger condition must be set to DATA using
TRIGger:A:BUS:B<x>:AUDio:CONDition.
B<x> is the bus number (1-4). 1-3 for the MDO4000C and 1-2 for the MDO3000.
2-784 MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual