Commands Listed in Alphabetical Order
HEADertime set
s the LIN error type to header time.
RESPtime sets the LIN error type to response time.
FRAMetime sets the LIN error type to frame time.
Examples
TRIGGER:A:BUS:B1:LIN:ERRTYPE CHECKSUM sets the LIN error type to
checksum.
TRIGGER:A:BUS:B1:LIN:ERRTYPE? might return
TRIGGER:A:BUS:B1:LIN:ERRTYPE SYNC indicating the LIN error type
is SYNC.
TRIGger:A:BUS:B<x>:LIN:IDentier:VALue
This command specifies the binary address string used for LIN trigger if the
trigger condition is ID or IDANDDATA.
B<x>
Conditions
This command requires a DPO4AUTO application module for DPO/MSO4000B
and MDO4000/B/C models or MDO3AUTO application module for
MDO3000 models.
Group
Trigger
Syntax
TRIGger:A:BUS:B<x>:LIN:IDentifier:VALu e <QString>
TRIGger:A:BUS:B<x>:LIN:IDentifier:VALu e?
Arguments
<QString> is the binary address string used for LIN trigger if the trigger
condition is ID or IDANDDATA.
Examples
TRIGGER:A:BUS:B1:LIN:IDENTIFIER:VALUE "110010" sets the identifier
value to 110010.
TRIGGER:A:BUS:B1:LIN:IDENTIFIER:VALUE? might return
TRIGGER:A:BUS:B1:LIN:IDENTIFIER:VALUE "XXXXXX" indicating the
identifier value is XXXXXX.
TRIGger:A:BUS:B<x>:MIL1553B:COMMAND:ADDRess:HIVALue
When the MIL-STD-1553 bus trigger condition is set to COMMAND,andthe
qualifier is set to
INrange or OUTrange , this command s
pecifies the upper
limit of the range for the remote terminal address field. (Use the command
2-830 MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual