Commands Listed in Alphabetical Order
Arguments
<NR3> is a floati
ng point number that specifies either the maximum remote
terminal response time (RT) or the inter-message gap (IMG) in seconds.
Examples
TRIGger:A:BUS:B1:MIL1553B:TIMe:MORELim it 80.0000e-6 would set
either the remote terminal response time (RT) or the inter-message gap (IMG) to
be used in the trigger to 80.0 μS.
TRIGger:A:BUS:B1:MIL1553B:TIMe:MORELim it? might return
100.0000e-6, indicating that the RT or IMG has been set to 100.00 μS.
TRIGger:A:BUS:B<x>:MIL1553B:TIMe:QUALifier
When the MIL-STD-1553 bus trigger condition is set to TIMe, this command
specifies the trigger data time qualifier. (This includes a smaller set of arguments
than other qualifier commands.)
B<x>
Conditions
This command requires a DPO4AERO application module for DPO/MSO4000B
and MDO4000/B/C models or MDO3AERO application module for
MDO3000 models.
Group
Trigger
Syntax
TRIGger:A:BUS:B<x>:MIL1553B:TIMe:QUALi fier
{LESSthan|MOREthan|INrange|OUTrange}
TRIGger:A:BUS:B<x>:MIL1553B:TIMe:QUALi fier?
Related Commands
TRIGger:A:BUS:B<x>:MIL1553B:CONDition
TRIGger:A:BUS:B<x>:MIL1553B:TIMe:LESSLimit
TRIGger:A:BUS:B<x>:MIL1553B:TIMe:MORELimit
Arguments
LESSthan sets the Time qualifier to less than minimum.
MOREthan sets the Time qualifier to greater than maximum
INrange sets the Time qualifier to inside range.
OUTrange sets the Time qualifier to out of range.
Examples
TRIGger:A:BUS:B1:MIL1553B:TIMe:QUALifi er LESSthan sets the Time
qualifier to less than minimum.
MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual 2-853