Commands Listed in Alphabetical Order
Group
Trigger
Syntax
TRIGger:A:BUS:B<x>:USB:SPLit:SE:VALue
{NOCARE|FULLSPeed|LOWSPeed|ISOSTART|IS OMID|ISOEND|ISOALL}
TRIGger:A:BUS:B<x>:USB:SPLit:SE:VALue?
Related Commands
TRIGger:A:BUS:B<x>:USB:SPECIALType
Arguments
NOCARE — any combination of S and E bits.
FULLSPeed —Sbit=0,Ebit=0.
LOWSPeed —Sbit=1,Ebit=0.
ISOSTART —Sbit=1,Ebit=0.
ISOMID — see note above.
ISOEND — see note above.
ISOALL — see note above.
Examples
TRIGger:A:BUS:B1:USB:SPLit:SE:VALue ISOEND specifies to trigger on
the ISOEND split transaction value.
TRIGger:A:BUS:B1:USB:SPLit:SE:VALue? might return FULLSPEED.
TRIGger:A:BUS:B<x>:USB:TOKENType
This command specifies the token type for the USB trigger. B<x>
Conditions
This command requires the DPO4USB application module to be installed, except
for MDO3000 models, which require the MDO3USB. Note that 1 GHz bandwidth
models are recommended for high-speed (HS) USB.
Group
Trigger
Syntax
TRIGger:A:BUS:B<x>:USB:TOKENType {ANY|SOF|OUT|I N|SETUP}
TRIGger:A:BUS:B<x>:USB:TOKENType?
Related Commands
TRIGger:A:BUS:B<x>:USB:CONDition
2-872 MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual