Commands Listed in Alphabetical Order
TRIGger:A:HO
LDoff? (Query Only)
Returns the A trigger holdoff parameters. These parameters specify the time
period during which the trigger circuitry is not looking to generate a trigger event.
Group
Trigger
Syntax
TRIGger:A:HOLDoff?
Related Commands
TRIGger:A:HOLDoff:TIMe
Examples
TRIGGER:A:HOLDOFF? might return :TRIGGER:A:HOLDOFF:TIME
900.0000E-09;BY DEFAULT
indicating that the A edge trigger holdoff time
(by default) is 900 ns.
TRIGger:A:HOLDoff:TIMe
This command specifies the A trigger holdoff time.
Group
Trigger
Syntax
TRIGger:A:HOLDoff:TIMe <NR3>
TRIGger:A:HOLDoff:TIMe?
Arguments
<NR3> is a floating point number that specifies the holdoff time in seconds. The
range is from 20 ns through 8.0 s.
Examples
TRIGGER:A:HOLDOFF:TIME ? might return :TRIGGER:A:HOLDOFFTIME
1.2000E-06
indicating that the A trigger holdoff time is set to 1.2 µs.
TRIGGER:A:HOLDOFF:TIME 10 sets the A trigger holdoff time to 10 s.
TRIGger:A:LEVel:AUXin
This command sets the threshold voltage level to use for an Edge, Pulse Width,
Runt or Rise/Fall (aka Transition, aka Slew Rate) trigger when triggering on the
Aux Input connector signal.
MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual 2-877

Table of Contents

Other manuals for Tektronix MDO3000 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix MDO3000 Series and is the answer not in the manual?

Tektronix MDO3000 Series Specifications

General IconGeneral
BrandTektronix
ModelMDO3000 Series
CategoryTest Equipment
LanguageEnglish

Related product manuals