EasyManuals Logo

Tektronix MDO3000 Series Program Manual

Tektronix MDO3000 Series
1158 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #937 background imageLoading...
Page #937 background image
Commands Listed in Alphabetical Order
TRIGger:A:TI
MEOut:TIMe
When triggering using the TIMEOut trigger type, this command specifies the
timeout time, in seconds. The default and minimum is 4.0E-9 seconds and the
maximum is 8.
0 seconds. The resolution is 800.0E-12 which means that the
increments of time specified is 800 picoseconds. The timeout trigger type is
selected using TRIGger:A:PULse:CLAss TIMEOut
Group
Trigger
Syntax
TRIGger:A:TIMEOut:TIMe <NR3>
TRIGger:A:TIMEOut:TIMe?
Related Commands
TRIGger:A:TIMEOut:POLarity
TRIGger:A:TIMEOut:SOUrce
TRIGger:A:PULse:CLAss
Arguments
<NR3> is a floating point number that specifies the timeout time, in seconds.
Examples
:TRIGger:A:TIMEOut:TIMe 4.0E-9 specifies the timeout time of 4.0 nsec.
:TRIGger:A:TIMEOut:TIMe? might return 8.0000E-9, which is the default.
TRIGger:A{:TRANsition|:RISEFall}? (Query Only)
Returns transition time trigger parameters.
Group
Trigger
Syntax
TRIGger:A{:TRANsition|:RISEFall}?
Related Commands
TRIGger:A:UPPerthreshold:CH<x> , TRIGger:A:LOWerthreshold:CH<x>
Examples
TRIGGER:A::TRANSITION? might return
:TRIGGER:A:TRANSITION:POLARITY POSITIVE;WHEN
SLOWER;DELTATIME 8.0000E-9
MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual 2-907

Table of Contents

Other manuals for Tektronix MDO3000 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix MDO3000 Series and is the answer not in the manual?

Tektronix MDO3000 Series Specifications

General IconGeneral
BrandTektronix
ModelMDO3000 Series
CategoryTest Equipment
LanguageEnglish

Related product manuals