Commands Listed in Alphabetical Order
TRIGger:A:TI
MEOut:TIMe
When triggering using the TIMEOut trigger type, this command specifies the
timeout time, in seconds. The default and minimum is 4.0E-9 seconds and the
maximum is 8.
0 seconds. The resolution is 800.0E-12 which means that the
increments of time specified is 800 picoseconds. The timeout trigger type is
selected using TRIGger:A:PULse:CLAss TIMEOut
Group
Trigger
Syntax
TRIGger:A:TIMEOut:TIMe <NR3>
TRIGger:A:TIMEOut:TIMe?
Related Commands
TRIGger:A:TIMEOut:POLarity
TRIGger:A:TIMEOut:SOUrce
TRIGger:A:PULse:CLAss
Arguments
<NR3> is a floating point number that specifies the timeout time, in seconds.
Examples
:TRIGger:A:TIMEOut:TIMe 4.0E-9 specifies the timeout time of 4.0 nsec.
:TRIGger:A:TIMEOut:TIMe? might return 8.0000E-9, which is the default.
TRIGger:A{:TRANsition|:RISEFall}? (Query Only)
Returns transition time trigger parameters.
Group
Trigger
Syntax
TRIGger:A{:TRANsition|:RISEFall}?
Related Commands
TRIGger:A:UPPerthreshold:CH<x> , TRIGger:A:LOWerthreshold:CH<x>
Examples
TRIGGER:A::TRANSITION? might return
:TRIGGER:A:TRANSITION:POLARITY POSITIVE;WHEN
SLOWER;DELTATIME 8.0000E-9