Commands Listed in Alphabetical Order
TRIGGER:B:EDG
E:SLOPE?
might return :TRIGGER:B:EDGE:SLO PE RISE
indicating that the B edge trigger occurs on the rising slope.
TRIGger:B:EDGE:SOUrce
This command specifies the source for the B trigger.
Group
Trigger
Syntax
TRIGger:B:EDGE:SOUrce {CH1|CH2|CH3|CH4|AUX|LINE |RF}
TRIGger:B:EDGE:SOUrce?
Related Commands
TRIGger:B:EDGE?
Arguments
CH1–4 specifies an analog channel as the B trigger source.
AUX specifies an external trigger using the Aux In connector located on the
front panel of the oscilloscope as the B trigger source. (For MSO/DPO 4000B
series models, MDO4000C series models without an RF option, and 2-channel
MDO3000 models only.)
LINE specifies the AC power line as the B trigger source.
RF specifies to use the internal RF power level as the source. (MDO4000/B/C
models only). Also requires installation of an MDO4TRIG application module.)
Examples
TRIGGER:B:EDGE:SOURCE CH4 sets channel 4 as the input source for the B
trigger.
TRIGGER:B:EDGE:SOURCE? might return : TRIGGER:B:EDGE:SOURCE CH1
indicating that the current input source for the B trigger is channel 1.
TRIGger:B:EVENTS? (Query Only)
Returns the current B trigger events parameter.
Group
Trigger
Syntax
TRIGger:B:EVENTS?
Related Commands
TRIGger:B:EVENTS:COUNt
MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual 2-923