Commands Listed in Alphabetical Order
Examples
TRIGGER:B:EVE
NTS?
might return
:TRIGGER:B:EVENTS:COUNT 2
indicating that 2 events must occur before the B trigger occurs.
TRIGger:B:EVENTS:COUNt
This command specifies the number of events that must occur before the B trigger
(when TRIG:DELay:BY is set to EVENTS).
Group
Trigger
Syntax
TRIGger:B:EVENTS:COUNt <NR1>
TRIGger:B:EVENTS:COUNt?
Related Commands
TRIGger:B:EVENTS?
Arguments
<NR1> is the number of B trigger events, which can range from 1 to 10,000,000.
Examples
TRIGGER:B:EVENTS:COUNT 4 specifies that the B trigger will occur four trigger
events after the A trigger.
TRIGGER:B:EVENTS:COUNT? might return :TRIGGER:B:EVENTS:COUNT 2
indicating that two events must occur after the A trigger before the B trigg
er
can occur.
TRIGger:B:LEVel
This co
mmand specifies the level for the B trigger.
Group
Trigger
Syntax
TRIGger:B:LEVel {TTL|<NR3>}
TRIGger:B:LEVel?
Related Commands
TRIGger:B, TRIGger:B:EDGE:SOUrce
Arguments
TTL specifies a preset TT L high level of 1.4 V.
<NR3> is a floating point number that specifies the B trigger level, in volts.
2-924 MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual

Table of Contents

Other manuals for Tektronix MDO3000 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix MDO3000 Series and is the answer not in the manual?

Tektronix MDO3000 Series Specifications

General IconGeneral
BrandTektronix
ModelMDO3000 Series
CategoryTest Equipment
LanguageEnglish

Related product manuals