Maintenance
Clearing the er
ror log. To clea r the error log after xing a problem, follow these
steps:
1. Enable the Ser
vice menus. (See page 5-2, Enable the Service Menu.)
2. Push the UTILITY button and push the Service option button to display
the Service
menu.
3. Push the Service Diag. option button to display the Diagnostics menu.
4. Push the Clear Error Log option button to clear the error log of past failures.
This also sets the Power Cycle count to zero.
TBS1000 Series Oscilloscope Service Manual 6–35