Reference
88
TDS 200-Series Digital Oscilloscope User Manual
Key Points
Self Calibration. The self calibration routine optimizes the oscilloscope
accuracy for the ambient temperature. For maximum accuracy,
perform a self cal if the ambient temperature changes by 5° C or
more.
To compensate the signal path, disconnect any probes or cables from
the input connectors. Then, select Do Self Cal to confirm that you
are ready to proceed.
System Status
Selecting System Status from the utility menu displays the menus
available for obtaining a list of control settings for each group of
instrument controls.
Push any front-panel menu button to remove the status screen.
Menu Settings Comments
Horizontal Lists horizontal parameters of channels
Vertical Lists vertical parameters of channels
Trigger Lists trigger parameters
Misc Lists the model of the instrument and the
software version number
If an extension module is installed, lists
values of the communications parameters