EasyManua.ls Logo

Tektronix TDS 420A - Page 287

Tektronix TDS 420A
433 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Command Descriptions
TDS Family Oscilloscope Programmer Manual
2–273
TRIGger:MAIn
Sets the main trigger level and returns the current main trigger parameters.
Trigger
TRIGger:MAIn SETLevel
TRIGger:MAIn?
TRIGger
:
MAIn
SETLevel
?
<Space>
SETLevel sets the main trigger level to half way between the MIN and MAX
amplitudes of the trigger source input. This is equivalent to pressing the
front-panel SET LEVel TO 50% button.
This argument works differently with AMI Communications Triggering settings.
Specifically, when AMI is chosen, this command measures the peak-to-peak
level and sets an upper threshold value at 75% and a lower threshold value at
25% of the peak-to-peak value. If the pulse form is CMI, NRZ, or an Eye
Diagram, the level is set to 50%.
TRIGGER:MAIN SETLEVEL
sets the main trigger level mid way between MAX and MIN.
TRIGger:MAIn:COMMunication:AMI:PULSEForm
TDS 500C & 700C Only
Sets or queries the communication trigger AMI pulse form to one of three
possibilities.
Trigger
TRIGger:MAIn:COMMunication:AMI:PULSEForm { PLUSOne | MINUSOne |
EYEdiagram }
TRIGger:MAIn:COMMunication:AMI:PULSEForm?
Group
Syntax
Arguments
Examples
Group
Syntax

Table of Contents

Other manuals for Tektronix TDS 420A