Command Descriptions
2−1 16
TDS200/1000/2000 Series Oscilloscope Programmer Manual
Measurement
MEASUrement:IMMed:VALue?
<NR3>
MEASUrement:IMMed:VALue?
might return 28.75E6 if you are measuring the frequency of a
28.75 MHz signal.
MEASUrement:IMMed:VALue?
might return 9.9E37, which means that the channel specified by
MEASUrement:IMMed:SOUrce is not currently displayed.
*ESR?
after a measurement query might return 16, indicating that there are
error messages in the queue.
ALLEv?
might return :ALLEV 541,Measurement warning, Low signal
amplitude: "
*ESR?, EVENT?, ALLEv?
MEASUrement:MEAS<x>? (Query Only)
Returns all measurement parameters for the displayed oscilloscope
periodic measurement specified by <x>, where <x> is 1 through 4
for the TDS200 series, and 1 through 5 for the TDS1000 and
TDS2000 series.
Measurement
Group
Syntax
Returns
Examples
Related Commands
Group

Table of Contents