Command Descriptions
TDS200/1000/2000 Series Oscilloscope Programmer Manual
2−1 17
MEASUrement:MEAS<x>?
Settings for the specified measurement source.
MEASUrement:MEAS3?
might return PERIOD;s";CH1
MEASUrement:MEAS<x>:SOUrce
Sets or queries the source for the measurement. This is equivalent to
selecting the measurement source in the MEASURE menu.
Measurement
MEASUrement:MEAS<x>:SOUrce CH<y>
MEASUrement:MEAS<x>:SOUrce?
CH<y> specifies the measurement source channel.
MEASUrement:MEAS2:SOUrce CH1
Sets the source of MEAS2 to channel 1.
MEASUrement:MEAS<x>:TYPe
Sets or queries the on-screen periodic oscilloscope measurement type
for the measurement specified by <x>. This is equivalent to selecting
the measurement type in the MEASURE menu. Setting the type to
anything other than NONE displays the MEASURE menu on the
screen.