Math FFT
The Math FFT function includes three FFT Window options. There is a trade-off
between
frequency resolution a nd amplitude accuracy with each type of window.
What you want to measure and your source signal characteristics will help you to
determine which window to use.
Window Measur
e
Charac
teristics
Hanning
Perio
dic waveforms
Bette
r frequency, poorer magnitude
accuracy than Flattop.
Flatt
op
Periodic waveforms Better magnitude, poorer frequency
accuracy than Hanning.
Rectangular Pulses or transients
Special-purpose window for w aveforms
tha
t do not have discontinuities. This is
essentially the same as no window.
FFT Aliasing
Problems occur when the oscilloscope acquires a time-domain waveform
containing frequency components that are greater than the Nyquist frequency.
(See page 63, Nyquist Frequency.) The frequency components that are above the
Nyquist frequency are undersampled, appearing as lower frequency components
t
hat "fold back" around the Nyquist frequency. These incorrect components are
called aliases.
66 TPS2000B Series Digital Oscilloscope User Manual

Table of Contents

Other manuals for Tektronix TPS2012B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix TPS2012B and is the answer not in the manual?

Tektronix TPS2012B Specifications

General IconGeneral
BrandTektronix
ModelTPS2012B
CategoryTest Equipment
LanguageEnglish

Related product manuals