6 Using the product
36
4. Enter values: [], [] and sometimes [], [] [OK].
- Particulate matter probe is preheated to 50 °C (status display
flashes green).
- Measurement system is operational (status display lights up
green).
5. [Next]
Adjustment help measurement: continue with point 9
Compliance measurement: continue with point 6
Approval test: continue with point 6
6. Start draught measurement: [].
- Draught zeroing.
7. Position the particulate matter probe in the centre of flow (area
of the highest flue gas temperature).
The display showing the maximum measured flue gas
temperature (FT max) helps when positioning the probe.
- The reading is displayed.
8. End draught measurement: []
9. [Next]
If the message Stability not yet achieved. Please wait…
appears, do not start any measurement until the message
is no longer displayed.