6 Using the product
37
10. Start measurement: [].
- The stability time (approx. 3 min) elapses. Following this, the
measuring phase begins automatically.
The measuring phase can be interrupted early:
> press [Cancel] .
- The measurement result is displayed once the measuring
phase has been completed
11. Prepare the instrument for rinsing.
> Remove particulate matter probe from the flue gas duct
[OK].
- Rinsing starts
12. Clean the sensor module, see Cleaning/replacing the sensor
module, page 64.
13. [OK].
> A test report can be printed out if necessary.
Data is printed out via the [] key or the Options menu.
Only readings assigned a display field in the measurement view
will be saved / printed out.
Measurement data can be printed out during a
measurement.
To be able to transfer data to a record printer via infrared or
Bluetooth interface, the printer used must be enabled.