9Error messages
Cause: The measuring rate selected with the pro-
gram cannot be reached with the given
probe constellation.
Remedy: - Check min. rate
- Remove any unnecessary probes
- Slow down measuring rate
- Restart
Cause: The measured value memory is either full
or blocked because a measuring protocol
is being displayed or printed.
Remedy: - Quit the running program
- Quit “M
EMORY
-VIEW
- Wait for printing or quit “P
RINT
Cause: The program running cannot be changed.
Remedy: Quit program and restart.
Cause: The entered program start criterion can-
not be fulfilled:
- Starting time in the past
- Competing trigger criteria on the same
measuring channel.
Remedy: Correct entries for correctness and
change if necessary e.g.
- incorrect (past) time marks
- Difference trigger when only one probe
is assigned.
Cause: No recognisable probes connected
Remedy: Connect probe to program measured
value trigger.
Rate does
not match
the sensor
connected
Program
cannot
be
executed
Program
running
Program-
ming
error
No
sensors
connected