8.3 Memory
8.3.3.2 Rate
Specify the rate at which the measuring instrument
should save the probe data during this measurement.
Entries from 0.1 seconds to 24 hours are possible.
Seconds can be given in fractions. Minutes and hour-
ly rates are usually whole numbers. All values are
saved according to the probes connected.
8Main menu
20
Program
Rate
˛1s
Ein
On
Enter /
Please note that some measuring time must be
allowed for each connected probe since the sockets
must be serially checked.
The probe with the longest measuring time
defines the minimum measuring rate.
The following is the time required for each probe:
Hum. probe °C/%RH w. dew-point .0.3 seconds
Vanes . . . . . . . . . . . . . . . . . . . . . . . .0.7 seconds
Shell anemometer . . . . . . . . . . . . . .2.2 seconds
Frequency sources . . . . . . . . . . . . .0.7 seconds
Pt100 probes . . . . . . . . . . . . . . . . .0.1 seconds
Other . . . . . . . . . . . . . . . . . . . . . . . .0.2 seconds
In the menu MIN.RATE (See Chap. 8.3.4) the short-
est measuring rate depending on the probes con-
nected can be checked.