7 Carrying out a measurement
17
7 Carrying out a measurement
7.1 Preparing for measurement
Prior to every measurement, please ensure that the instrument is in perfect
condion:
• For example, keep an eye out for broken housing or leaking baeries.
• Always carry out a funcon test before using the instrument, see below.
• Check that the instrument is funconing perfectly (for example at a known
voltage source) before and aer every test.
• If the safety of the user cannot be guaranteed, the instrument must be
switched o and secured to prevent unintenonal usage.
When connecng the test leads to the test object, always connect the
common test lead (COM) to the test object rst of all. When
disconnecng the test leads, always disconnect the +/- phase test lead
rst of all.
Serious risk of injury to the user and/or destrucon of the instrument while
measuring current.
> Measuring circuit must be de-energized.
The measuring instrument may only be used in circuits up to a
maximum voltage of of 600 V (testo 770-1/-2/-3 (0590 7703)) /
1000 V (testo 770-3 (0590 3770)). The nominal cross-secon of the
connecon cable must be taken into account in order to ensure safe
connecon (e.g. via crocodile clips).
Strong RF interference and / or open leads when measuring A AC may
result in unstable display readings.
For the A DC and μA DC measurement always zero the instrument with
[ ] before measuring.