9 Technical data
27
9.2 More technical data
9.2.1 testo 770-1/-2
4.000 V
40.00 V
400.0 V
600 V
± (0.8 % of m.v. + 3 digits)
4.000 V
40.00 V
400.0 V
600 V
± (1.0 % of m.v. + 3 digits)
DC current
- jaws [A]
- jack [µA]
(testo 770-2)
± (2.0 % of m.v. + 5 digit)
± (2.0 % of m.v. + 5 digits)
± (1.5 % of m.v. + 5 digits)
AC current
3
- jaws [A]
5
- jack [µA]
(testo 770-2)
2
,
4
± (2.0 % of m.v. + 5 digit)
± (2.0 % of m.v. + 5 digits)
± (1.5 % of m.v. + 5 digits)
400.0 Ω
4.000 kΩ
40.00 kΩ
400.0 kΩ
4.000 MΩ
40.00 MΩ
0.1 Ω
1 Ω
10 Ω
100 Ω
1 kΩ
10 kΩ
± (1.5 % of m.v. + 3 digits)
± (1.5 % of m.v. + 5 digits)
The lower measuring ranges are only specied from 5% (does not apply to DC current / AC
current measurements with the current probe)
Frequency of AC voltage 40 Hz to 1 kHz
In the case of a mixed signal (AC + DC), only the purely AC component is taken into account
As the frequency increases (over 400 Hz), the accuracy deteriorates
+/- (2.5% of m.v. + 3 digits) for 400Hz to 750Hz / +/- (5.0% of m.v. + 3digits) for 750Hz to 1000Hz
Frequency of AC current up to 400 Hz
Specicaon is valid for capacitances > 10 nF