9 Technical data
27
9.2 More technical data
9.2.1 testo 770-1/-2
Feature
Measuring
range
1
Resolution
Accuracy
DC voltage
4.000 V
40.00 V
400.0 V
600 V
1 mV
10 mV
100 mV
1 V
± (0.8 % of m.v. + 3 digits)
AC voltage
2
,
3
,
4
4.000 V
40.00 V
400.0 V
600 V
1 mV
10 mV
100 mV
1 V
± (1.0 % of m.v. + 3 digits)
DC current
- jaws [A]
- jack [µA]
(testo 770-2)
40 A
400 A
400 µA
0.1 A
0.1 A
0.1 µA
± (2.0 % of m.v. + 5 digit)
± (2.0 % of m.v. + 5 digits)
± (1.5 % of m.v. + 5 digits)
AC current
3
- jaws [A]
5
- jack [µA]
(testo 770-2)
2
,
4
40 A
400 A
400 µA
0.1 A
0.1 A
0.1 µA
± (2.0 % of m.v. + 5 digit)
± (2.0 % of m.v. + 5 digits)
± (1.5 % of m.v. + 5 digits)
Resistance
400.0 Ω
4.000
40.00
400.0
4.000
40.00
0.1 Ω
1 Ω
10 Ω
100 Ω
1
10
± (1.5 % of m.v. + 3 digits)
Connuity alarm
<0 30 Ω
Diode test
yes (0 to 2.5 V)
Capacity
51.20 nF
6
0.01 nF
± 10 % typically
512.0 nF
0.01 nF
± (1.5 % of m.v. + 5 digits)
1
The lower measuring ranges are only specied from 5% (does not apply to DC current / AC
current measurements with the current probe)
2
Frequency of AC voltage 40 Hz to 1 kHz
3
In the case of a mixed signal (AC + DC), only the purely AC component is taken into account
4
As the frequency increases (over 400 Hz), the accuracy deteriorates
+/- (2.5% of m.v. + 3 digits) for 400Hz to 750Hz / +/- (5.0% of m.v. + 3digits) for 750Hz to 1000Hz
5
Frequency of AC current up to 400 Hz
6
Specicaon is valid for capacitances > 10 nF

Table of Contents