8 Maintenance
40
8.4 Firmware update
The current firmware version is available at www.testo.com.
There are two options available:
Performing an update using IRSoft or
Performing an update with the thermal imager
Downloading firmware
1
- Download firmware: Firmware-testo-865-872.exe
2
- Unzip file: double-click on the exe. file
FW_T87x_Vx.xx.bin is stored in the selected folder.
8.4.1 Performing an update using IRSoft
8.4.1.1 Preparing the camera
- The rechargeable battery is fully charged or the camera is connected
to the mains adapter.
1
- Connect the computer to the camera with the USB cable.
2
- Switch on the camera.
8.4.1.2 Performing an update
IRSoft is activated.
1
- Select Camera Configuration.
2
- Click on Configure thermal imager.
The Camera settings window opens.
3
- Select Instrument settings Firmware Update OK.
FW_T87x_Vx.xx.bin is displayed.
4
- Select Open.
Firmware update is performed.
Thermal imager is automatically restarted. Update complete.