Heracell 150i/240i 50115191_C_01-2018 177 Thermo Scientific
Chapter 13 Data communication
13.6.2. Overview of possible event entries in bit coding
Code Event Special information (byte 8-15)
0x01 Set values from all control loops (periodi-
cally in minute cycles)
Current values for temperature, CO
2
, O
2
and rh
0x02 Set value change (at beginning of new
section)
Set value for temperature, CO
2
, O
2
and rh
0x10 Change temperature set value Set value for temperature, CO
2
, O
2
and rh
0x11 Change CO
2
set value Set value for temperature, CO
2
, O
2
and rh
0x12 Change O
2
set value Set value for temperature, CO
2
, O
2
and rh
0x20 New temperature error Status/error registration for temperature,
CO
2
, O
2
and rh
0x21 New CO
2
error Status/error registration for temperature,
CO
2
, O
2
and rh
0x22 New O
2
error Status/error registration for temperature,
CO
2
, O
2
and rh
0x2F New system error Status/error registration for temperature,
CO
2
, O
2
and rh
0x30 Power reset Set value for temperature, CO
2
, O
2
and rh
0x31 Door open Current actual values for temperature,
CO
2
, O
2
and rh
0x32 Door closed Current actual values for temperature,
CO
2
, O
2
and rh
0x40 Customer-specific temperature calibration Calibration level (2 bytes), old temperature,
new temperature (2 bytes each)
0x41 Customer specific CO
2
calibration Calibration level (2 bytes), old temperature,
new temperature (2 bytes each)
0x42 Customer specific O
2
calibration Calibration level (2 bytes), old temperature,
new temperature (2 bytes each)
0x50 Start auto-start Status/error registration for temperature,
CO
2
, O
2
and rh
0x51 auto-start successfully completed Current actual values for temperature,
CO
2
, O
2
and rh
0x52 auto-start completed with error Status/error registration for temperature,
CO
2
, O
2
and rh
0x53 auto start stopped manually Status/error registration for temperature,
CO
2
, O
2
and rh
0x60 Start contra-con Status/error registration for temperature,
CO
2
, O
2
and rh
0x61 contra-con successfully completed Current actual values for temperature,
CO
2
, O
2
and rh
0x62 contra-con completed with error Status/error registration for temperature,
CO
2
, O
2
and rh
0x63 contra-con stopped manually Status/error registration for temperature,
CO
2
, O
2
and rh