HyPerforma DS 300 SingleUse Mixer User's Guide | 45
Table 4.4. Minimum open‑top mixing for adjustable motor head S.U.M.s with standard
impeller sizes.
Left-
to-right
offset
Front-
to-back
position
Lift height
Motor angle
(fixed or
adjustable
motor head
models)
Motor
rpm
Minimum
working
liquid
volume
Turn-
down
ratio
Minimum
hold-up
Maximum
liquid fill
50 L
drum
≈ 1 in. left ≈ 1/2 in. 0 5° 180 9.2 L 5.4:1 100 mL 50 L
100 L
drum
≈ 1 in. left ≈ 1/2 in. 0 5° ≈ 205 9.8 L 10:1 100 mL 100 L
100 L
drum
≈ 1 in. left ≈ 1/2 in. ≈ 7/8 in. 5° 140 16.7 L 6:1 100 mL 100 L
200 L
drum
≈ 3/4 in.
left
≈ 1 1/4 in. ≈ 11 1/16 in. 5° 150 20 L 10:1 100 mL 200 L
200 L
drum
≈ 3/4 in.
left
≈ 1 1/4 in. ≈ 11 1/8 in. 5° 140 22.5 L 8.8:1 100 mL 200 L
300 L
drum
NA due to
flat bottom
NA ≈ 15.24 in. 5° 164 61 L 5:1 500 mL 300 L
300 L
drum
NA due to
flat bottom
NA ≈ 15.24 in. 5° 124 65 L 5:1 500 mL 300 L
Table 4.5. Country‑specific electrical plug part numbers.
Country Cat. no.
Europe SV50177.345
UK SV50177.346
China SV50177.347
US (110 VAC) SV50177.348
US (208 VAC) SV50177.349
India SV50177.350
Italy, Chile SV50177.351
Brazil, Switzerland SV50177.352