B-36 Micro-Tech 9101/9201 Reference Manual, Rev J Thermo Fisher Scientific
5D Current output #4
should be ____
4.00 mA (4-20) (0-20) (20-0)
(20-4)
“
#2 -0.00 V
This scroll is only available when
an analog I/O board is installed.
It shows the status of each
#4 -0.00 V
A
*Only available when COMM
board is installed.
A test pattern is sent out on the
TX output and read on the RX
input. If the test fails, the
message Test Failed is shown;
otherwise, the message TEST
PASSED is displayed.
(Password: Service.)
B
This scroll only displays when the
Profibus board is installed.
10
12
13 Keyboard test Press keys. Hit RUN key twice
quickly to exit.
The USB menu allows you to download all the Micro-Tech settings to a
flash drive to back them up and, in the event of a system failure, easily
restore them from the flash drive. You can also print to the flash drive.
Be sure to only use the USB stick that is provided with the Micro-Tech.
If using another USB stick with the Micro-Tech, please note it must have
a capacity of 1GB or less, and must be properly formatted as described
below before being used with the Micro-Tech.
Due to limitations in the FTDI VNC1L USB Controller IC, the USB
devices used with the Micro-Tech must be formatted with 512-byte data
sectors. The default sector-size applied when a USB device is formatted
in Microsoft Windows, is not necessarily 512 bytes, so a special
procedure is required.