C-22 Micro-Tech 9101/9201 Reference Manual, Rev J Thermo Fisher Scientific
Command Word
Command
Value
Description Action Condition
Int
Bit
Hex
16 4 10
Remote Acq
Calibration (16)
At the end of the calibration functions (to
determine the end check the “Calibration
Running” in the Status 1 register), after
verifiying the Error (Calibration Error
registers in Block 0), this command tells
the Micro-Tech to save the new Zero or
new Span (according to which calibration
function has been performed).
If this command is sent during a remote
calibration, it aborts the running function.
speed signal
8 3 8 Batch Standby (8)
Temporarily stops a Batch (Load Out)
sequence. Batch can resume if a Start
command is sent later.
Batch Enabled
4 2 4 Batch Abort (4)
Stops the Batch (Load Out) sequence.
Batch Enabled
2 1 2 Batch Start (2)
Start the Batch (Load Out) sequence.
Batch Enabled
1 0 1
Acknowledge Alarms
(1)
Reset the pending alarms if these have
been previously acknowledged.
Alarm Active
This section provides details about the Siemens 3964R protocol.
Two communication modes are provided, as follows
FETCH
The Host asks the slave unit for registers.
SEND
The Host sends registers to the slave.
The choice between requesting or sending data is made using a specific
code inside the message. Here are four examples.
Siemens
3964R