Thermo Fisher Scientific Micro-Tech 9101/9201 Reference Manual, Rev J C-5
The following procedure applies to the Host.
Sends data to the Micro-Tech.
Waits at least 100 ms.
Reads the success flag. It should be 0.
Some data is read only, some is read/write, and some is write only.
NOTE. In all cases, the maximum number of words the system can
transfer is limited to 41 per time. Requests of registers in excess of 41 are
considered errors and do not generate an answer.
When a COMM line is set up for printing, the system prints out data in
different ways, depending on the set-up. Print menus become visible, which
allow the user to define when and in which format the data is printed.
Periodical printouts at predefined intervals, as well as at predefined
times of day, or on command are possible.
The format of the printouts is selectable between a number of
predefined formats, plus a fully programmable user-defined format. In
this particular case, the user is able to easily format the printout to fit
into any pre-printed ticket or form, without the need of special
software.