Daily Operation
Things to Do After Operating the LTQ
@@@@@@@@@@@@@@@@@@@@@@@@@@@@@@@@@@@@@@@@@@@@
'JOOJHBO-52
5IFSNP
&-&$530/$03103"5*0/
3-10 ______________________ Finnigan LTQ Hardware Manual _______________________
i. Place the ion sweep cone and the ion capillary tube in a beaker of
50:50 methanol/water.
ii. Sonicate it for 15 min.
iii. Dry the ion sweep cone.
d. Clean the spray cone with a Kimwip soaked in methanol.
e. Reinstall the ion sweep cone as follows:
i. Note the location of the sweep gas supply port in the API cone
seal. The gas inlet on the ion sweep cone is placed in this port.
See Figure 3-1 and Figure 3-2.
Figure 3-1. Sweep gas supply port in the API cone seal
Sweep
Gas
Supply
Port
Spray Cone