System Shutdown, Startup, and Reset
Resetting the Data System
@@@@@@@@@@@@@@@@@@@@@@@@@@@@@@@@@@@@@@@@@@@@@@@@@@@@
'JOOJHBO-52
5IFSNP
&-&$530/$03103"5*0/
5-12 ______________________ Finnigan LTQ Hardware Manual _______________________
3. Press <Ctrl>-<Alt>-<Del> when you are prompted to do so. Then, click
on OK or enter your password (if you have one) in the Logon Information
dialog box to complete the shutdown and restart procedure.
4. When the shutdown and restart procedure has completed, choose
Start > Programs > Xcalibur > LTQ Tune to display the Tune Plus
window.
Note. The communications link between the data system and the MS
detector should be automatically reestablished after you reset the data
system. When this occurs the Communication LED on the front panel of the
MS detector is illuminated yellow and then green. If the system is unable to
reestablish the communications link, press the Reset button (for 3 s) on the
power panel of the MS detector.