Orion Benchtop pH/ISE Meter Instruction Manual
Overview of Meter Functions-420Aplus,710Aplus
99
Code Function Description
6-2 DATALOG Sets the timed datalogger interval. The
INTERVAL maximum interval is Interval 23 hours, 59
minutes, and 59 seconds. The minimum
interval is five (5) seconds. When the code
6-2 is displayed in the lower field, the current
datalogger interval hours are displayed in
the main field indicated by an H in the left
position. Press or to change the setting,
then press yes. For the current datalogger
interval minutes: seconds will be displayed in
the main field. Press or to change, then
press yes. The default setting is (5) seconds.
6-3 TRANSFER Transfers data from the datalogger to a
DATA printer or a computer. When the code 6-3 is
displayed the OFF will flash. To transfer the
data, press or until ON is displayed,
then press yes. Once the data has been
transferred, the meter advances to the next
setup option.
NOTE: Downloading the data does not clear
the memory. To clear memory, see step 6-4 below.
6-4 CLEAR Clears all data from the datalogger memory.
DATA When the code 6-4 is displayed, the OFF will
flash. To clear the data, press or until ON
is displayed. The alarm will ring rapidly to
indicate the memory is about to be erased.
Press yes.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Thermo Orion Aplus Series and is the answer not in the manual?

Thermo Orion Aplus Series Specifications

General IconGeneral
BrandThermo
ModelOrion Aplus Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals