Orion Benchtop pH/ISE Meter Instruction Manual
Overview of Meter Functions 520Aplus,525Aplus,720Aplus and 920Aplus
108
Key Decimal Point Data/print modes
Press the decimal point to set the datalog function and the print mode. See
the descriptions below for instructions. Print mode is presented first, then
the datalog function.
Printing
Press data/print modes to select a print option, then follow the prompts:
Prompt Keypress/Action
Press data/print modes.
1. MANUAL PRINT The current print mode is displayed. Wait for
the prompt.
2. SET PRINT MODE? Press yes.
3. 1-RDY The three print options are displayed
2-TIM along with current setting:
3-MAN 1 For Print on ready.
2 For Print at a timed interval.
3 For manual Print.
Enter 1, 2, or 3 and press yes.
4. MANUAL PRINT Your choice is verified, then the meter
proceeds to steps 6-10 if timed print selected.
5. PRT INTRVL HRS Enter the print interval hours setting, 00 to
23, then press yes.
6. PRT INTRVL MIN Enter the print interval minutes setting, 00 to
59, then press yes.
7. PRT INTRVL SEC Enter the print interval seconds setting, 00 to
59, then press yes.
8. PRINT INTERVAL The print interval is verified then the meter
continues.
9. SET LOG MODE? Press yes to continue (see below) or no
to exit.