Page 31
PM103x Optical Power Meter Interfaces Chapter 5 Specifications
5.2 PM103E
All technical data are valid at 23 ± 5°C and 45 ± 15% rel. humidity (non condensing).
Item #
PM103E
Sensor Compatibility
Compatible Sensor Types
Thorlabs' C-Series Photodiode and Pyroelectric Sensors
Photodiodes (Max 10 mA)
Pyroelectric Detectors (Max 200 V, pulse only)
Auxiliary Input / Output
Digital I/O
Phoenix DMC connector 0,5/7
Interfaces
Ethernet
RJ45
RS232
Phoenix DMC connector 0,5/7
Photodiode Sensor: Current Input
Connector
Female DE-9
Units
W, dBm, W/cm², A
Measurement Range
21 Ranges:
nA: 2, 4, 10, 20, 40, 100, 200, 400;
µA: 1, 2, 4, 10, 20, 40, 100, 200, 400;
mA: 1, 2, 4, 10
Range Selectable [W], Sensor Dependent
Display Resolution
1 pA / Responsivity Value (A/W)
Measurement Uncertainty
±0.2% Full Scale (100 nA - 10 mA)
±0.5% Full Scale (2 nA - 40 nA)
Analog Bandwidth
DC - 100 kHz, Dependent on Sensor and Settings
Wavelength Correction
Sensor Dependent; nm, (V/W)
Beam Area Setting
Diameter 1/e² or Rectangular x,y
AD Converter
16 bit
Pyroelectric Sensors: Pulsed Voltage Input
Connector
Female DE-9
Units
J, J/c, W, dBm, W/c, V
Measurement Range
13 Ranges;
mV: 20, 40, 100, 200, 400;
V: 1, 2, 4, 10, 20, 40, 100, 200;
Ranges Selectable in J, Sensor Dependent
Display Resolution
1 µV / Responsivity Value (V/W)
Measurement Uncertainty
±0.5% Full Scale (100 mV - 200 V)
±1% Full Scale (20 mV, 40 mV)
Trigger Level
3 % - 90 % Full Scale
Repetition Rate
Single Shot - 80 kHz, Dependent on Sensor
Wavelength Correction
nm, (V/W) (Sensor Dependent)
Beam Area Setting
Diameter 1/e² or Rectangular x,y
AD Converter
16 bit
Sensor Temperature Measurement
Supported Temperature Sensor
Thermistor

Other manuals for THORLABS PM103