Page #28 background imageLoading...
Page #28 background image
© 2018 Thorlabs26
TSP01
3. *OPC? – operation complete query – places a “1” in output queue
When *OPC is sent, the OPC bit in the Standard Event Register will set after all pending com-
mand operations are complete. When *OPC? is sent, an ASCII “1” is placed in the Output
Queue after all pending command operations are complete.
Typically, either one of these commands is sent after the INITiate command. The INITiate
command is used to take the instrument out of idle in order to perform measurements. While
operating within the trigger model layers, many sent commands will not execute. After all pro-
grammed operations are completed, the instrument returns to the idle state at which time all
pending commands (including *OPC and/or *OPC?) are executed. After the last pending com-
mand is executed, the OPC bit and/or an ASCII “1” is placed in the Output Queue.
4. *RST – reset – return instrument to defaults
When the *RST command is sent, the instrument performs the following operations:
· Cancels all pending commands.
· Cancels response to any previously received *OPC and *OPC? commands.
5. *TST? – self-test query – run self test and read result
Use this query command to perform the instrument self-test routine. The command places the
coded result in the Output Queue. A returned value of zero (0) indicates that the test passed,
other values indicate that the test failed.
6. *WAI – wait-to-continue – wait until previous commands are completed
The *WAI command is a no operation command for the instrument and thus, does not need to
be used. It is there for conformance to IEEE488.2.
4.2.2.2 SCPI Command Reference
SYSTem subsystem commands
Command
Description
SCPI
SYSTem
Path to SYSTem subsystem
R
:ERRor
R
[:NEXT]?
Returns the latest error code and message
R
:VERSion?
Returns level of SCPI standard (1999.0)
R
STATus subsystem commands
Command
Description
SCPI
STATus
R
:OPERation
Path to control operation event registers
R
[:EVENt]?
Returns the event register
R
:CONDition?
Returns the condition register
R
:ENABle <value>
Sets the enable register
R
:ENABle?
Returns the enable register
R
:QUEStionable
Path to control questionable event registers
R
[:EVENt]?
Returns the event register
R
:CONDition?
Returns the condition register
R
:ENABle <value>
Sets the enable register
R
:ENABle?
Returns the enable register
R
:PRESet
Set status registers to default states.
R

Other manuals for THORLABS TSP01

Question and Answer IconNeed help?

Do you have a question about the THORLABS TSP01 and is the answer not in the manual?

THORLABS TSP01 Specifications

General IconGeneral
TypeTemperature and Humidity Sensor
DisplayLCD
InterfaceUSB
Temperature Measurement Range-40 to 100 °C
Temperature Measurement Accuracy±0.5 °C
Humidity Measurement Range0 to 100% RH
Dew Point Measurement Range-40 °C to 100 °C
Dew Point Measurement Accuracy±1 °C
Sampling Rate1 Hz
Power SupplyUSB Powered
Operating System CompatibilityWindows
Humidity Measurement Accuracy±2 %RH (10 %RH to 90 %RH at 25 °C)

Summary

General Information

Ordering Codes and Accessories

Lists available product codes and accessories for the TSP01.

Getting Started

Parts List

Lists all items included in the TSP01 package.

System Requirements

Specifies the minimum hardware and software requirements for the TSP01.

Software Installation

Guides the user through the TSP01 application software installation process.

Operating Instructions

Quick Start Guide

Provides a brief overview of using the TSP01 application.

Detailed GUI Description

Explains the functions of the TSP01 graphical user interface elements.

Write Your Own Application

TSP01 RevB Driver Installation

Details the driver installation and file locations for TSP01 RevB sensors.

Previous-Revision TSP01 Driver Installation

Details driver installation and file locations for older TSP01 sensors.

Command Reference

Details SCPI and IEEE488.2 commands for instrument control.

Troubleshooting

TSP01 Not Recognized Troubleshooting

Solutions for when the TSP01 is not recognized by the software.

Maintenance and Service

Version Information

Explains how to find software version information via the Help menu.

Appendix

Technical Data

Lists the detailed technical specifications of the TSP01 sensor.

Safety Precautions

Outlines important safety guidelines for operating the TSP01.

Thorlabs Worldwide Contacts

Lists contact information for Thorlabs support and sales globally.

Related product manuals