12. Dry paraffin filter and metal ring with a soft,
clean, dry cloth.
13. Replace paraffin filter. Replace metal ring (place
on top of filter).
WARNING: RETURN XYLENE AND
XYLENE-DAMPENED CLOTH TO SAFE
STORAGE PRIOR TO RESTORING
POWER TO THE MODULE.
NOTE: To add paraffin, refer to “Adding Paraffin” in
Section 3 of this manual.
CLEAN MAGNIFYING LENS
ASSEMBLY (OBTAINED BY SEPARATE
PURCHASE)
The magnifying glass and assembly may be
cleaned:
Q
while still attached to the Embedding Module
or
Q
when totally removed from Embedding
Module.
To remove non-paraffin debris:
1. Dampen a soft cloth with household glass cleaner.
2. Wipe magnifying glass and entire assembly to
remove fingerprints, dust particles and other non-
paraffin debris.
3. Dry entire assembly using a soft, clean, dry cloth.
To remove paraffin debris:
1. Dampen a soft cloth with xylene.
2. Wipe assembly to remove paraffin and other debris.
2. WARNING: XYLENE IS FLAMMABLE. BE
SURE THE MODULE IS COOL PRIOR TO
USING XYLENE.
2. CAUTION: Do not immerse assembly in
xylene or frame surrounding lens may soften
or disintegrate.
3. Dry assembly using a soft, clean, dry cloth.
Cryo Module
Clean Exterior Surfaces/Ventilation Panels
1. WARNING: SET POWER SWITCH ON
FRONT PANEL OF CRYO MODULE TO
OFF. UNPLUG POWER CORD FROM
OUTLET.
2. Using a clean cloth dampened in water or mild
soap solution, remove dust/debris from exteri-
or surfaces, ventilation panels and work area
(surfaces under and around module).
3. Rinse as necessary. Dry using a soft, clean, dry
cloth.
4. Connect power cord to module. Plug power
cord into outlet.