CA-200F Corneal Analyser - Rev. 7 16/01/2012
2
NOTES
Installation with external devices
The CA-200F corneal analyser complies with the CE marking requirements.
Before connecting an external device (computer, printer, display, keyboard, mouse, slit lamp), check that
these devices are in compliance with the EN 60950-1 standards and bear the CE marking.
If the CA-200F corneal analyser is installed in rooms for medical use, any connected personal computer and
printer must be powered by an insulated transformer in accordance with IEC 60601-1.
If the CA-200F corneal analyser is installed in rooms for medical use and is not connected to a computer,
there is no need to use an insulated transformer. If a printer is connected directly to the CA-200F corneal
analyser, it must be powered by an insulated transformer in accordance with IEC 60601-1.
In the event of an accident or near accident due to use of the instrument, please contact:
Manufacturer
VISIA imaging S.r.l.
Via C. E. Gadda, 15
52027 San Giovanni Valdarno (AR)
Italy
In the event of a fault or to request technical service, contact:
Distributor
TOPCON EUROPE MEDICAL B.V.
Essebaan 11
2908 LJ Cappelle A/D Ijseel
THE NETHERLANDS
WARNING
The CA-200F device must be used with the power supply provided in the package
and in any case only with an EXM 80 5118 power supply for medical use.