CA-200F Corneal Analyser - Rev. 7 16/01/2012
6
The CA-200F device is intended for use in the electromagnetic environment specified below. The
customer or the user of the CA-200F device should assure that it is used in the said environment.
Electromagnetic environment ā
guidance
Electrostatic discharge
(ESD)
EN 61000-4-2
Floors should be made of wood,
concrete or ceramic tiles. If floors
are covered with synthetic
material, the relative humidity
should be at least 30%
Electrical fast transient/
burst
EN 61000-4-4
2kV for power
supply lines
2kV for
power supply lines
The quality of the power mains
should be that of a typical
commercial or hospital
environment.
The quality of the power mains
should be that of a typical
commercial or hospital
environment.
Voltage dips, short
interruptions and
voltage variations on
power supply input lines
EN 61000-4-11
< 5% U
T
(>95% dip in U
T
)
for 0.5 cycle
40% U T
(60% dip in U
T
)
for 5 cycle
70% U T
(30% dip in U
T
)
for 25 cycle
< 5% U T
(>95% dip in U
T
)
for 5 seconds
< 5% U T
(>95% dip in U
T
)
for 0.5 cycle
40% U
T
(60% dip in U
T
)
for 5 cycle
70% U T
(30% dip in U
T
)
for 25 cycle
< 5% U T
(>95% dip in U
T
)
for 5 seconds
The quality of the power mains
should be that of a typical
commercial or hospital
environment. If the user requires
continuous device operation
during power mains interruptions,
it is recommended that the device
be powered using an
uninterruptible power supply or a
battery.
Power frequency
magnetic field
EN 61000-4-8
Power frequency magnetic fields
should be at levels characteristic
of a typical location in a typical
commercial or hospital
environment.