− 7 −
R-CL300-0-1204-02
1-7-11 PD UNIT
The PD function is set up in this selection box.
ON → When selected, “PD” image is added in the MENU screen and the use of PD value display
can be selected.
OFF → When selected, the use of the PD value display cannot be selected.
1-7-12 PRINT CLEAR
ON → When selected, after PRINT operation is completed (including a print error), “CLEAR”
processing is performed.
OFF → When selected, after PRINT operation is completed (including a print error), “CLEAR”
processing is not performed.
1-7-13 STD1 FORMAT
1 → When selected, the PD measured value is not added to the STD1 format.
2 → When selected, the PD measured value is added to the STD1 format.
1-7-14 SC ARRANGE
ON → When selected, the SC value of “0.034” is replaced to “0.00”.
OFF → When selected, the value is displayed as it is.
1-7-15 MARK COLOR
The MARK color can be changed with this function.
PINK: When selected, the mark color is changed to pink.
ORANGE: When selected, the mark color is changed to orange.
1-7-16 CONTINUE
The target update frequency is changed with this function. Normally, “OFF” is selected.
ON: When selected, the update is performed once per second.
OFF: When selected, the update is performed 60 times per second.
1-7-17 LAN UNIT
ON: LAN connection can be selected.
OFF: LAN connection cannot be selected.
1-7-18 PROGRAM DL
This function is used when the program is downloaded.
1-7-19 PRINT CUT
ON: When selected, “Cut” is performed after PRINT operation is completed.
OFF: When selected, “Cut” is not performed after PRINT operation is completed.